ProLab 3000 Ion concentration
135
ba75645e03 12/2012
14 Start the measurement with [Continue].
The measured value is checked for stability (Stability control).
15 Wait for the measurement with stability control to end.
or
Take over the measured value with [OK].
The Calculation register with the calculated measured value
pops up.
16 Change the displayed unit of the measured value with [Unit] if
necessary.
17 If required, display the record of the current measurement with
[Record].
18 Start measuring further samples with [Continue].
Repeat steps 5 - 15 for all samples.
or
Terminate the measuring method with [Cancel] and confirm the
safety query with [OK].
Measurement with the selected method is completed.