Ion concentration ProLab 3000
114
ba75645e03 12/2012
12 In the Standard register, enter the values for Standard volume
and Standard concentration if necessary.
13 Use [Dosing] to automatically add to the sample the adjusted
volume of the standard solution (Standard volume)
or
If the burette control is inactive:
Add to the sample the adjusted volume of the standard solution
manually (Standard volume)
14 Start the measurement with [Continue].
The measured value is checked for stability (Stability control).
15 Wait for the measurement with stability control to end.
or
Take over the calibration value with [OK].
The Calculation register with the calculated measured value
pops up.