Xylem CellOx 325 Dissolved oxygen sensor Manuel d'utilisateur Page 15

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 16
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 14
CellOx 325 Technical data
31
Pin assignment
Measurement
conditions
Measuring ranges at 20 °C 0 ... 50 mg/l O
2
0 ... 600 % O
2
saturation
0 ... 1250 mbar O
2
partial pressure
Temperature range 0 ... 50 °C
Max. allowed overpressure 6·10
5
Pa (6 bar)
Depth of immersion min. 6 cm
max. 20 m (depending on the cable length)
Operating position any
Approach flow > 3 cm/s at 10 % measurement accuracy
10 cm/s at 5 % measurement accuracy
18 cm/s at 1 % measurement accuracy
Characteristic data
on delivery
Zero signal < 0.1 % of the saturation value
Response time at 20 °Ct
90
(90 % of the final value display) < 10 s
t
95
(95 % of the final value display) < 16 s
t
99
(99 % of the final value display) < 60 s
Oxygen own consumption
at 20 °C
0.008
µg·h
-1
(mg/l)
-1
Drift approx. 3 % per month in the operating con-
dition
Working life min. 6 months with one electrolyte fill
Vue de la page 14
1 2 ... 10 11 12 13 14 15 16

Commentaires sur ces manuels

Pas de commentaire