ba75373e04 07/2006VARIO pHOperating manualHand-held pH meter
Commissioning VARIO pH10ba75373e04 07/20063.2 Initial commissioningPerform the following activities:z Connect the electrode to the meter (see section
VARIO pH Commissioning11ba75373e04 07/2006If you do not wish to measure immediately, put the watering cap on the short pH electrode (see section 5.4).
Commissioning VARIO pH12ba75373e04 07/20063.4 Connecting an electrodeThe adapters make the meter compatible with commercial electrodes. You can turn t
VARIO pH Commissioning13ba75373e04 07/2006VARIO pH withconnectedelectrodeDIN adapterElectrodePlug of the NTC 30temperature sensorh:min1CLRLoBatDay.Mon
Commissioning VARIO pH14ba75373e04 07/20063.4.2 Electrode without temperature sensorConnect an electrode without an integrated temperature sensor to t
VARIO pH Commissioning15ba75373e04 07/20063.4.3 Electrode with S7 connectorElectrodes with a plug head connector can be connected directly to the S7 a
Commissioning VARIO pH16ba75373e04 07/20063.4.4 Connecting an external temperature probeWhen using an electrode without temperature sensor it is possi
VARIO pH Operation17ba75373e04 07/20064 Operation4.1 Operating elementsThe entire operation is carried out via the touch sensitive surface of the disp
Operation VARIO pH18ba75373e04 07/2006Touch screenoperationIn this operating manual, keys are indicated by angle brackets <..>. The touch scree
VARIO pH Operation19ba75373e04 07/20064.2 Switching on the meterNoteThe meter has an energy saving feature to avoid unnecessary battery depletion. The
VARIO pHba75373e04 07/2006Accuracy whengoing to pressThe use of advanced technology and the high quality standard of our instruments are the result of
Operation VARIO pH20ba75373e04 07/20064.3.2 Measuring the pH value4.3.3 Measuring the temperatureFor reproducible pH measurements, it is essential to
VARIO pH Operation21ba75373e04 07/2006NoteThe meter automatically recognizes a temperature sensor. The display accuracy of the temperature indicates w
Operation VARIO pH22ba75373e04 07/20064.4 CalibrationWhy calibrate? pH electrodes age. This changes the asymmetry (zero point) and slope of the pH ele
VARIO pH Operation23ba75373e04 07/2006NoteIf the slope (SLO) is indicated on the display, you can change the unit of the slope with <"> [&l
Operation VARIO pH24ba75373e04 07/20064.4.1 Carrying out calibrationPreparatoryactivitiesManualtemperature inputWhen measuring via an adapter with a s
VARIO pH Operation25ba75373e04 07/2006NoteAt this point, calibration can be terminated with <E>. This corresponds to a single-point calibration.
Operation VARIO pH26ba75373e04 07/2006Three-pointcalibrationNoteYou can also terminate the calibration prematurely with <E>. The values of the t
VARIO pH Operation27ba75373e04 07/20064.4.2 Displaying the calibration dataThe calibration data are displayed after calibrating (see section 4.4.1) an
Operation VARIO pH28ba75373e04 07/20064.4.3 Setting the calibration intervalThe calibration interval determines how many days after the last calibrati
VARIO pH Operation29ba75373e04 07/20064.5 SavingThe VARIO pH meter has a data memory. It can store up to 50 measurement datasets, depending on the sho
VARIO pH31 Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52 Safety . . . . . . . . . . . . . . . . . . . . .
Operation VARIO pH30ba75373e04 07/2006If all memory locations are occupied, the StoFull message appears. You have the following options:4.5.2 Displayi
VARIO pH Operation31ba75373e04 07/20064.5.3 Erasing measured data from the memoryTo erase all datasets proceed as follows:4.6 ClockThe meter has a clo
Operation VARIO pH32ba75373e04 07/2006Setting the dateand time1 Switch on the meter with <(>. After a short display test, a measured value is di
VARIO pH Operation33ba75373e04 07/20064.6.2 Alarm functionWith the alarm function you set a point of time (alarm time) at which an alarm signal should
Operation VARIO pH34ba75373e04 07/2006Switching off thealarm function4.6.3 StopwatchWith the stopwatch, you measure time periods of up to 59:59 (minut
VARIO pH Operation35ba75373e04 07/20064.6.4 TimerThe timer function can count down the seconds of a specified time interval (timer interval) of maximu
Operation VARIO pH36ba75373e04 07/20064.7 Switch-off intervalThe automatic switch-off switches off the meter automatically if no entry is made for a s
VARIO pH Operation37ba75373e04 07/20064.8 Turning the displayWhen operating the meter with the adapter, the display indication can be turned by 180 de
Operation VARIO pH38ba75373e04 07/20064.9 ResetThe following measurement parameters are reset to the delivery status:NoteThe calibration data gets los
VARIO pH Maintenance, cleaning, disposal39ba75373e04 07/20065 Maintenance, cleaning, disposal5.1 MaintenanceThe meter is almost maintenance-free. The
VARIO pH46 What to do if... . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 437 Technical data . . . . . . . . . . . . . . . . . .
Maintenance, cleaning, disposal VARIO pH40ba75373e04 07/20065.3 CleaningOccasionally wipe the outside of the meter with a damp, soft cloth.CautionThe
VARIO pH Maintenance, cleaning, disposal41ba75373e04 07/20065.4 StorageTo maintain the operability of the meter and short pH electrode for a long time
Maintenance, cleaning, disposal VARIO pH42ba75373e04 07/20065.5 TransportActivate the cleaning mode. Thus, the instrument will not interpret any conta
VARIO pH What to do if...43ba75373e04 07/20066 What to do if...Error message0FL0FL0FL0FL, UFLUFLUFLUFL Error messageErr3Err3Err3Err3 Cause Remedy–
What to do if... VARIO pH44ba75373e04 07/2006Error messageErr4Err4Err4Err4Error messageErr5Err5Err5Err5No stablemeasured value Cause RemedyNo short
VARIO pH What to do if...45ba75373e04 07/2006Display of l Instrument doesnot reactto keystroke Sensor symbolflashes Message StoFullStoFullStoFullStoFu
What to do if... VARIO pH46ba75373e04 07/2006
VARIO pH Technical data47ba75373e04 07/20067 Technical dataDepending on the connected sensor (short pH electrode or external electrode), the VARIO pH
Technical data VARIO pH48ba75373e04 07/20067.2 Measuring ranges, resolution, accuracy7.2.1 Short pH electrode SenTix V7.2.2 Adapter with electrode (DI
VARIO pH Lists49ba75373e04 07/20068ListsThis chapter provides additional information and orientation aids.Abbreviations The list of abbreviations expl
VARIO pH Overview5ba75373e04 07/20061 OverviewThe hand-held VARIO pH meter enables you to measure the pH value of a solution very simply. All function
Lists VARIO pH50ba75373e04 07/2006AbbreviationsASY AsymmetryNIST pH calibration with buffer solutions produced according to DIN 19266TEC pH calibratio
VARIO pH Lists51ba75373e04 07/2006GlossaryAdjusting To manipulate a measuring system so that the relevant value (e. g. the displayed value) differs as
Lists VARIO pH52ba75373e04 07/2006pH value The pH is a measure of the acidic or basic effect of an aqueous solution. It corresponds to the negative de
VARIO pH Index539 IndexAAlarm function ...33Asymmetry ...22Authorized
Index VARIO pH54
Wissenschaftlich-Technische Werkstätten GmbHDr.-Karl-Slevogt-Straße 1D-82362 WeilheimGermanyTel: +49 (0) 881 183-0+49 (0) 881 183-100Fax: +49 (0) 881
Overview VARIO pH6ba75373e04 07/2006NoteCertain sensors enhance the scope of functions of the VARIO pH. More detailed information on the presently ava
VARIO pH Safety7ba75373e04 07/20062SafetyThis operating manual contains basic instructions that you must follow during the commissioning, operation an
Safety VARIO pH8ba75373e04 07/20062.2 General safety instructionsThis instrument is built and inspected according to the relevant guidelines and norms
VARIO pH Commissioning9ba75373e04 07/20063 Commissioning3.1 Scope of deliveryVARIO pH singlemeterz VARIO pH hand-held meterz WTW Technical buffers– pH