ba75689e01 01/2007
Multi 3500i
Operating manual
ESC
STO
P
R
T
C
AL
M
O
pH / ISE / DO / conductivity measuring instrument
Vue de la page 0
1 2 3 4 5 6 ... 106 107

Résumé du contenu

Page 1 - Multi 3500i

ba75689e01 01/2007Multi 3500iOperating manualESCSTOPRTCALMOpH / ISE / DO / conductivity measuring instrument

Page 3 - Multi 3500i Contents

Lists Multi 3500i98GlossaryAdjusting To manipulate a measuring system so that the relevant value (e. g. the displayed value) differs as little as poss

Page 4 - Contents Multi 3500i

Multi 3500i Lists99Measured value The measured value is the special value of a measured parameter to be determined. It is given as a combination of th

Page 5 - 1Overview

Lists Multi 3500i100Salinity The absolute salinity SA of seawater corresponds to the relationship of the mass of dissolved salts to the mass of the so

Page 6 - 1.2 SETs of equipment

Multi 3500i Index1019 IndexAAir calibration vessel ...53Asymmetry of the pH electrode ...30Authorized use .

Page 7 - Multi 3500i Overview

Index Multi 3500i102MultiLab pilot ... 77OObligations of the purchaser ... 10Operating safety

Page 8 - 1.3 Keypad

Multi 3500i Appendix: Firmware Update103Appendix: Firmware UpdateGeneral The program "Firmware Update 350i" enables you to update the firm-w

Page 9 - 1.5 Socket field

Appendix: Firmware Update Multi 3500i104

Page 10 - Overview Multi 3500i

Wissenschaftlich-Technische Werkstätten GmbHDr.-Karl-Slevogt-Straße 1D-82362 WeilheimGermanyTel: +49 (0) 881 183-0+49 (0) 881 183-100Fax: +49 (0) 881

Page 11 - 2 Safety

Multi 3500i Safety92 SafetyThis operating manual contains basic instructions that you must follow during the commissioning, operation and maintenance

Page 12 - Safety Multi 3500i

Safety Multi 3500i102.2 General safety instructionsThis instrument is built and inspected according to the relevant guide-lines and norms for electron

Page 13 - 3 Commissioning

Multi 3500i Commissioning113 Commissioning3.1 Scope of deliveryz Multi 3500i handheld measuring instrument with 4 rechargeable bat-teries, 1.2 V type

Page 14 - 3.3 Initial commissioning

Commissioning Multi 3500i123.3 Initial commissioningPerform the following activities:z For mains operation and charging the batteries: Connect the plu

Page 15 - 4 Operation

Multi 3500i Operation134 Operation4.1 Switching on the measuring instrumentSwitching on Press the <ON/OFF> key.The measured value display appear

Page 16 - Operation Multi 3500i

Operation Multi 3500i144.2 General operating principlesThis section contains basic information of the operation of the Multi 3500i. Operating elements

Page 17 - Multi 3500i Operation

Multi 3500i Operation154.2.2 NavigationMeasured value display In the measured value display, you can z select a measuring window with <▲><▼&g

Page 18

Operation Multi 3500i16z FunctionsFunctions are designated by the name of the function. They are im-mediately carried out by confirming with <MENU/

Page 19

Multi 3500i Operation174.2.3 Example 1 on navigation: Setting the language1 Press the <ON/OFF> key.The measured value display appears.The instru

Page 20

Accuracy whengoing to pressThe use of advanced technology and the high quality standard of our instruments are the result of a continuous development.

Page 21 - Date format: dd.mm.yy

Operation Multi 3500i186 Open the General submenu with <MENU/ENTER>. 7 Open the setting mode for the Language with <MENU/ENTER>.8 Select

Page 22

Multi 3500i Operation194.2.4 Example 2 on navigation: Setting the date and timeThe measuring instrument has a clock with a date function. The date and

Page 23

Operation Multi 3500i205 Change and confirm the setting with <▲><▼> and <MENU/ENTER>.The hours are highlighted.6 Change and confirm

Page 24

Multi 3500i Operation214.3 System settings (system menu)The system menu comprises the following settings:z Data storage (see section 4.3.1)z Display (

Page 25

Operation Multi 3500i22Settings To open the system menu in the measured value display, press the <MENU/ENTER> key for approx. 1 s. After complet

Page 26

Multi 3500i Operation234.3.3 SystemOverview The following sensor-independent instrument features can be adjusted in the system menu/System and its sub

Page 27 - 4.4 pH value / ORP voltage

Operation Multi 3500i24System / Interface / Baud rate1200, 2400, 4800, 9600, 19200Baud rate of the data inter-faceSystem / Interface / Output formatAS

Page 28

Multi 3500i Operation254.4 pH value / ORP voltage4.4.1 General informationYou can measure the following variables:z pH value [ ]z ORP [mV]CautionWhen

Page 29

Operation Multi 3500i26The display of the temperature indicates the active temperature mea-suring mode:* If you do not wish that, you can:– either dis

Page 30

Multi 3500i Operation274.4.2 Measuring the pH valueAutoRead(Drift control)The AutoRead function (drift control) continually checks the stability of th

Page 31

Multi 3500i Contents11 Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31.1 General features . . . . . . . . .

Page 32

Operation Multi 3500i284.4.3 Measuring the ORP voltageNoteORP electrodes are not calibrated. However, you can check ORP elec-trodes using a test solut

Page 33

Multi 3500i Operation294.4.4 Settings for pH and ORP measurementsOverview The following settings are possible for pH and ORP measurements:z Resolution

Page 34

Operation Multi 3500i304.4.5 pH calibrationWhy calibrate? pH electrodes age. This changes the asymmetry (zero point) and slope of the pH electrode. As

Page 35

Multi 3500i Operation31Calibration points Calibration can be performed using one, two or three buffer solutions in any order (single-point, two-point

Page 36 - Continue

Operation Multi 3500i32Display calibration dataand output to interfaceYou can have the data of the last calibration displayed (see section 4.8.5). Sub

Page 37

Multi 3500i Operation33Calibration evaluation After calibrating, the measuring instrument automatically evaluates the calibration. The asymmetry and s

Page 38

Operation Multi 3500i344.4.6 Carrying out the TEC and NIST/DIN calibrationThe two calibration procedures only differ in the usage of different buff-er

Page 39

Multi 3500i Operation35NoteFor single-point calibration, the instrument uses the Nernst slope (-59.16 mV/pH at 25 °C) and determines the asymmetry of

Page 40 - Set buffer: 7.00

Operation Multi 3500i36Continuing for two-pointcalibration(Calibration type TEC)8 Thoroughly rinse the electrode with distilled water.9 Immerse the el

Page 41

Multi 3500i Operation37Continuing for three-point calibration(Calibration type TEC)14 Thoroughly rinse the electrode with distilled water.15 Immerse t

Page 42

Contents Multi 3500i24.7 Conductivity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 584.7.1 General information . . . . . . .

Page 43 - 4.5 Ion concentration

Operation Multi 3500i384.4.7 Carrying out a ConCal calibrationSingle-point calibration Use any buffer solution for this rapid method.The calibration w

Page 44

Multi 3500i Operation39NoteFor single-point calibration, the instrument uses the Nernst slope (-59.16 mV/pH at 25 °C) and determines the asymmetry of

Page 45

Operation Multi 3500i40Continuing for two-pointcalibration(Calibration typeConCal)10 Thoroughly rinse the electrode with distilled water.11 Immerse th

Page 46

Multi 3500i Operation414.5 Ion concentration4.5.1 General informationNoteIncorrect calibration of ion sensitive electrodes will result in incorrect me

Page 47

Operation Multi 3500i42Preparatory activities Perform the following preparatory activities when you want to measure:NoteWhile no valid calibration is

Page 48 - ISE <CAL>

Multi 3500i Operation43adjust the temperature of the standard and measuring solutions ac-cordingly. If the temperature difference is greater the [Temp

Page 49

Operation Multi 3500i44Standard solutions Use two or three different standard solutions. For three-point calibra-tion, the standard solutions have to

Page 50

Multi 3500i Operation45NoteThe calibration record is automatically transmitted to the interface after calibrating.Sample record:Calibration evaluation

Page 51 - 4.6 Dissolved oxygen

Operation Multi 3500i46Preparatory activities Perform the following preparatory activities when you want to calibrate:Carrying out anISE calibrationPr

Page 52

Multi 3500i Operation47Continuing for two-pointcalibration11 Wait for the end of the AutoRead measurement or accept the calibration value with <MEN

Page 53

Multi 3500i Overview31Overview1.1 General featuresThe Multi 3500i compact precision handheld meter enables you to car-ry out pH measurements, ISE meas

Page 54

Operation Multi 3500i48Continuing for three-point calibrationRepeat the steps 12 to 20 in the same way for the third standard solu-tion. After finishi

Page 55

Multi 3500i Operation494.6 Dissolved oxygen4.6.1 General informationYou can measure the following variables:z DO concentration z DO saturation index (

Page 56

Operation Multi 3500i50Preparatory activities Perform the following preparatory activities when you want to measure:4.6.2 MeasuringSelecting thedispla

Page 57

Multi 3500i Operation51NoteThe ConOx double sensor can perform the salinity correction automat-ically. The conductivity module of the sensor measures

Page 58

Operation Multi 3500i52Settings The settings are made in the measuring menu of the DO measurement. To open the settings, activate the relevant measuri

Page 59

Multi 3500i Operation534.6.4 DO calibrationWhy calibrate? DO sensors age. This changes the slope of the DO sensor. Calibration determines the current

Page 60 - 4.7 Conductivity

Operation Multi 3500i54Sample record:Calibration evaluation After the calibration, the measuring instrument evaluates the current status of the sensor

Page 61

Multi 3500i Operation55Calibration inwater vaporsaturated air(air calibration vessel)For this calibration procedure, the Comparison meas. setting must

Page 62

Operation Multi 3500i56Calibration via aComparison meas.For this calibration procedure, the Comparison meas. setting must be set to On in the measurin

Page 63

Multi 3500i Operation575 Press <MENU/ENTER>.The AutoRead measurement to determine the DO concentrati-on starts.6 Wait for the end of the AutoRea

Page 64

Overview Multi 3500i41.2 SETs of equipmentThe measuring instrument is also available as part of individual SETs of equipment. You will find additional

Page 65

Operation Multi 3500i584.7 Conductivity4.7.1 General informationYou can measure the following variables:z Conductivityz Specific resistancez Salinityz

Page 66 - <CAL>

Multi 3500i Operation594.7.2 MeasuringYou can carry out conductivity measurements as follows:Selecting thedisplayedmeasured variableYou can switch bet

Page 67

Operation Multi 3500i604.7.3 Temperature compensationThe calculation of the temperature compensation is based on the pre-set reference temperature, 20

Page 68 - 4.8 Data storage

Multi 3500i Operation614.7.4 Settings for conductivity measuring cellsOverview The following settings are possible for conductivity measuring cells:z

Page 69

Operation Multi 3500i62Temp. comp. (TC) / Linear coeff.0.000 ... 3.000 %/KCoefficient of the linear tem-perature compensation.This setting is only ava

Page 70

Multi 3500i Operation634.7.5 Determining the cell constant (calibration in the control standard)Why determine the cellconstant?Aging slightly changes

Page 71

Operation Multi 3500i64Calibration evaluation After the calibration, the measuring instrument automatically evaluates the current status of the calibr

Page 72 - Stop automatic storage?

Multi 3500i Operation655 Press <MENU/ENTER>.The AutoRead measurement to determine the cell constant starts.6 Wait for the end of the AutoRead me

Page 73

Operation Multi 3500i664.8 Data storageYou can transmit measured values (data records) to the data storage in two ways:z Manual data storage (see sect

Page 74 - - TDS 689 mg/l 25.0 °CAR

Multi 3500i Operation674.8.1 Manual data storageProceed as follows to transmit to the data data storage and simulta-neously output to the interface a

Page 75

Multi 3500i Overview5Set (sample configuration):1 Multi 3500i measuring instrument, carrying strap with 2 carrying clips, armoring2 Cond/Oxi beaker wi

Page 76

Operation Multi 3500i684.8.2 Automatic data storage at intervalsThe storage interval (Interval) determines the chronological interval be-tween automat

Page 77 - Sample printout

Multi 3500i Operation69Data storage adminis-trationThe two lower display lines indicate the use of the data storage calcu-lated in advance for the sel

Page 78

Operation Multi 3500i70Terminating the auto-matic data storage func-tion prematurelyProceed as follows to switch off the automatic data storage functi

Page 79

Multi 3500i Operation71SettingsMenu item Setting/functionDescriptionData storage / Manual data storage / Display- Displays in pages all mea-suring dat

Page 80 - 4.10 Reset

Operation Multi 3500i72Data filterDisplay presentationof a datasetMenu item Setting/functionDescriptionFilter Filter criteria:No filter Data filter sw

Page 81

Multi 3500i Operation73Sample printoutQuitting the display To quit the display of stored measuring datasets, you have the follow-ing options:z Switch

Page 82

Operation Multi 3500i744.8.5 Displaying and outputting calibration recordsYou can display the calibration data z of one selected sensor, orz of all se

Page 83 - 5.3 Disposal

Multi 3500i Operation75Sample printout03.11.03 07:14CALIBRATION pH03.11.03 07:10:45Multi 3500i ser. no. 12345678Cal. interval 7 dAutoCal TE

Page 84

Operation Multi 3500i764.9 Transmitting data (RS 232 interface)4.9.1 Options for data transmissionVia the RS 232 interface, you can transmit data to a

Page 85 - 6 What to do if

Multi 3500i Operation77interface (displayed measured values, measuring datasets, calibration records).4.9.2 Connecting a PC/external printerUse the AK

Page 86 - What to do if... Multi 3500i

Overview Multi 3500i61.3 KeypadKey functionsMENUENTERESCSTOCALPRTMSelect the measured variable <M>:– pH value / ORP / ion concentration– DO conc

Page 87 - ± 2 °C temperature

Operation Multi 3500i784.10 ResetYou can reset (initialize) all sensor settings and sensor-independent settings separately from each other.4.10.1 Rese

Page 88 - 6.3 DO measurement

Multi 3500i Operation79Cond The following settings for conductivity measurements are reset to the default settings with the Reset function:The sensor

Page 90 - 6.5 General errors

Multi 3500i Maintenance, cleaning, disposal815 Maintenance, cleaning, disposal5.1 MaintenanceThe measuring instrument is maintenance-free.5.2 Cleaning

Page 91 - 7 Technical data

Maintenance, cleaning, disposal Multi 3500i82Measuring instrument Dispose of the measuring instrument without the rechargeable batter-ies as electroni

Page 92 - Technical data Multi 3500i

Multi 3500i What to do if...836 What to do if...6.1 pH and ORP measurementError message,OFL or UFL(measuring rangeexceeded or undercut)Error message,

Page 93 - Multi 3500i Technical data

What to do if... Multi 3500i84No stable measuredvalue Obviously incorrectmeasured valuesSensor symbol flashesCause RemedypH electrode:– Diaphragm con

Page 94

Multi 3500i What to do if...856.2 ISE measurementError message OFLObviously incorrectmeasured valuesError message, Error(invalidcalibration) Warning [

Page 95

What to do if... Multi 3500i866.3 DO measurementDO sensor was notrecognizedError message OFLError message, Error(invalid calibration)Error message, LE

Page 96

Multi 3500i What to do if...87Measured value toohighSensor symbol flashes6.4 Conductivity measurementConductivitymeasuring cellwas not recognizedError

Page 97

Multi 3500i Overview71.4 DisplayThe graphical display can indicate up to three measuring windows at the same time. The illumination enables to read th

Page 98

What to do if... Multi 3500i886.5 General errorsDisplay, LoBatInstrument does notreact to keystrokeCause Remedy– Batteries almost empty – Charge the b

Page 99 - U Voltage

Multi 3500i Technical data897 Technical data7.1 General dataDimensions approx. 172 x 80 x 37 mmWeight approx. 0.3 kg (without plug-in power supply)Mec

Page 100 - Lists Multi 3500i

Technical data Multi 3500i90Guidelinesand norms usedEMC EC guideline 89/336/EECEN 61326-1:1998EN 61000-3-2 A14:2000EN 61000-3-3:1995FCC Class AInstrum

Page 101 - Multi 3500i Lists

Multi 3500i Technical data917.2 Measuring ranges, resolution, accuracy7.2.1 pH/ORP7.2.2 ISEMeasuring ranges,resolutionVariable Measuring range Resolut

Page 102

Technical data Multi 3500i927.2.3 Dissolved oxygenMeasuring ranges,resolutionNote: The values quoted in brackets apply especially for the DurOx 325 se

Page 103 - Multi 3500i Index

Multi 3500i Technical data937.2.4 ConductivityMeasuring ranges,resolutionVariable Measuring range Resolution [µS/cm] 0.000 ... 2.000*0.00 ... 20.00**0

Page 104 - Index Multi 3500i

Technical data Multi 3500i94Accuracy (± 1 digit) Variable Accuracy Temperature of the test sample / Temperature compensationNone (Off) ± 0.5 %Nonline

Page 105 - Appendix: Firmware Update

Multi 3500i Lists958ListsThis chapter provides additional information and orientation aids.Abbreviations The list of abbreviations explains the indica

Page 106

Lists Multi 3500i96AbbreviationsConductivity value (international γ)ASY AsymmetryCalibration type NIST/DINAutomatic pH calibration with buffer solutio

Page 107

Multi 3500i Lists97TDS Total Dissolved SolidsTr20 Reference temperature of 20 °CTr25 Reference temperature of 25 °CUFL Display range undercut (underfl