Operation Multi 3400i
60
4.9 Reset
You can reset (initialize) the measurement parameters and the config-
uration parameters separately from one another.
Measurement
parameters
The following measuring parameters (pH/O
2
/Cond InI) can be reset to
the default condition:
pH
O
2
Cond
Note
The calibration data gets lost when the measuring parameters are re-
set. Recalibrate after performing a reset.
Configuration
parameters
The following configuration parameters (InI) are reset to the delivery
status:
Measuring mode pH
Asymmetry 0 mV
Slope -59.16 mV
Measuring mode D. O. concentration
Relative slope 1.00
Salinity (value) 0.0
Salinity (function) Off
Measuring mode
Cell constant 0.475 cm
-1
(fixed)
Baud rate 4800
Display during the
pH calibration
Buffer nominal value
Interval 1
(automatically saved) OFF
Interval 2
(for data transmission) OFF