Global Water
OWNER’S MANUAL
CL500 OnLine Residual
Chlorine Monitor
Global Water Instrumentation
11390 Amalgam Way
Gold River CA 95670
Phone: 800-876-1172
Fax: 847-672-9988
Website:
www.globalw.com
Catalog #24420GW (2/08)
REV. 2.0
Vue de la page 0
1 2 3 4 5 6 ... 35 36

Résumé du contenu

Page 1

Global Water OWNER’S MANUAL CL500 OnLine Residual Chlorine Monitor Global Water Instrumentation 11390 Ama

Page 2

4.0 Installation and Commissioning Prior to use for the first time, one of the reagents (the indicator) will have to be mixed. Refer to section 10.2

Page 3 - Table of Contents

4.2 Plumbing The recommended plumbing for the instrument is shown in Figure 4. The instrument is designed to require very little head pressure to ope

Page 4

The fluid waste from drain connection of this instrument contains reagents diluted with large quantities of sample water. Global Water Instrumentation

Page 5 - Specifications

Plugs are inserted into the RS-485 and 4-20mA cable bulkheads when shipped, to ensure a watertight seal. These plugs should be removed and discarded w

Page 6 - 1.0 Overview

CL500 (2/08) Page 10 REV 2.0 4.3.3 Relays The Alarm 1 and Alarm 2 relays are mechanical relays rated at 240 VAC 2A. Please note that the relays are

Page 7

5.0 Operation The CL500 Online Chlorine Analyzer allows for the measurement of the chlorine of process water on-line. The chlorine value of the proce

Page 8

5.2 Security Access Feature The instrument is equipped with a security access code feature that can be activated in the configuration mode. If the

Page 9 - 3.0 Theory of Operation

6.0 Instrument Calibration The instrument was tested prior to leaving the factory. Since it operates from a pre-determined calibration curve, no cali

Page 10

6.2 Zero (offset) Calibration Procedure Generally this calibration is only required if readings are expected to be below 1 mg/L or if it is required

Page 11

7.0 Instrument Configuration (CONFIG mode)The instrument has been designed to provide the ability to customize the instrument according to needs at

Page 13

The next prompt will be the chlorine level assigned to the 20MA. Select the chlorine level using the S and T buttons. Once the desired level has been

Page 14

Press the  button to continue on and select the desired instrument address using the S or T buttons. Once the selection is satisfactory, press the 

Page 15 - 5.0 Operation

Alarm 1 Set Point: This prompt is used to select the set point for this alarm; this is indicated by “S/P” shown on the lower row of the display. Sele

Page 16 - button

7.7 Speed of Response The speed of response for both displayed and output values of mg/L can be adjusted in this menu. Although the default setting i

Page 17

7.9 RS-485 Parameters These menus will only appear if the RS-485 is enabled (see 7.3). The default is 8 Bit, no (nOnE) Parity, 1 Stop Bit. Make selec

Page 18

CL500 (2/08) Page 21 REV 2.0 8.0 Additional Features and Options 8.1 Backlit LCD The backlit LCD allows for easier readability of the LCD dis

Page 19 - ↵ button

CL500 (2/08) Page 22 REV 2.0 8.2.3 Modbus Communication Modbus protocol communication manual is available Cat. #24569GW. This manual is available

Page 20

CL500 (2/08) Page 23 REV 2.0 9.0 Troubleshooting 9.1 CL500 Fault Detection The CL500 performs continuous diagnostic monitoring. In the CL500, t

Page 21

CL500 (2/08) Page 24 REV 2.0 A level 1 fault is a system fault. This is NOT a problem that the operator can correct, and the unit must be returned

Page 22

CL500 (2/08) Page 25 REV 2.0 9.4 Diagnostic Chart Symptom Cause Cure Lower display shows MA 4-20 mA loop open Check wiring. See sections 4.3.4 a

Page 23

Table of Contents Section Page Specifications...

Page 24

10.0 Routine Maintenance 10.1 Maintenance Schedule The recommended schedule is shown below. It is important to replace the reagents on a monthly

Page 25

After a PRIME the CL500 will perform a water calibration (WCAL). It will take a few minutes to complete this procedure. 3 Month Tubing Replacemen

Page 26

INLETCHECKVALVESCAP ASSEMBLY "A"PUMPHAMMERAND SPRINGTHUMBSCREWREAGENTPUMPASSYINLETOUTLETFLOW`CHECK VALVE DETAILOPTICS SYSTEMKNURLED TOPC

Page 27 - 9.0 Troubleshooting

Indicator Reagent Preparation Add about 400 ml of deionized water into the indicator bottle. Cap tightly and shake vigorously until the powder is diss

Page 28

CL500 (2/08) Page 30 REV 2.0 10.5 Cleaning the CL500 Flush the system as mentioned in section 10.1 Maintenance Schedule. When the flushing is fin

Page 29 - Symptom Cause Cure

CL500 (2/08) Page 31 REV 2.0 11.0 Accessories and Replacement Parts List The items shown below are recommended accessories and replacement parts. A

Page 30 - 10.0 Routine Maintenance

CL500 (2/08) Page 32 REV 2.0 12.0 Warranty Global Water Instrumentation., as vendor, warrants to the original purchaser of this instrument that it

Page 31

Table of Contents (continued) Section Page 7.7 Speed of Response...

Page 32 - and the indicator

CL500 (2/08) Page 1 REV 2.0 Specifications Measurement Range 0 – 10.00 mg/L (PPM) Accuracy ±5% of reading or ±0.03 mg/L (PPM) whichever is greater f

Page 33

CL500 (2/08) Page 2 REV 2.0 1.0 Overview The CL500 Online Chlorine Analyzer allows for the reading of chlorine levels of process water on-line. The

Page 34

1.2 The Display Figure 1 illustrates all the items that can appear on the display. The upper row of the display (1) is used for reporting the chlor

Page 35 - Accessory Catalog Number

2.0 Safety This manual contains basic instructions that must be followed during the commissioning, operation, care and maintenance of the instrument

Page 36 - 12.0 Warranty

CL500 (2/08) Page 5 REV 2.0 3.0 Theory of Operation The CL500 has two solenoid valves, one for sample water (FLOW) and one for draining of the cuve