Uploading new FW to the Display unit
Background:
After uploading the FW to the Display unit, the language file must be uploaded again and
the preferred language selected in the “Display Setup” menu.
1. Current MagFlux® log data must be saved in a file before transferring and installing new
firmware. See "2. Save Log Data" overleaf.
2. Note down all display and converter settings before upgrading new firmware.
3. The upgrade procedures below must be carried out in the following order A - B - C.
A. Connect a PC to the Flow Meter
1. Insert the MagFlux® MJK-Field Link CD-ROM into the PC's disk drive.
The MJK-Field Link opening menu will appear after a short while (see below). If the auto-
run function does not execute, locate the "MJK-Field-Link 840110-0XX.exe" file (or similar)
and click "Run".
Vue de la page 3